Code 3 System Alert

Code 3

Process Overtemperature

Support Level
Tier 2 (Not User Resolvable)
System Analysis

Description

The temperature during the sterilization process has exceeded the allowed limit.

This error may occur when sterilization cycles are started one immediately after another, without sufficient time for the device to cool down. It may also occur if a cycle is run without a tray in the chamber.

Other possible causes include a malfunctioning electromagnetic valve, a faulty chamber temperature sensor, or a mainboard error.