System Diagnostics

Select a code below to view quick details, or access the full resolution guide.

Vacuum Test Failed System Alert

Vacuum Test Failed

Unable to perform the vacuum test

System Analysis

Description

The device was unable to perform the vacuum test.

Possible Causes:

  • A leak in the chamber
  • Leaky solenoid valves (V1 / V2 / V3)
  • A faulty safety valve
  • Issues with the steam generator or pressure sensor
  • A worn-out or dirty chamber door gasket, which prevents proper sealing.

1. Quick Fix

Clean the gasket with warm water, using cloth.

2. Replace the gasket if needed

If the gasket appears worn, deformed, or cracked — replace it.